LEADTOOLS Medical (Leadtools.Dicom assembly) Send comments on this topic. | Back to Introduction - All Topics | Help Version 17.0.3.29
WaveformGroupCount Property
See Also 
Leadtools.Dicom Namespace > DicomDataSet Class : WaveformGroupCount Property



Gets the number of waveform multiplex groups that are included in the dataset.

Syntax

Visual Basic (Declaration) 
Public ReadOnly Property WaveformGroupCount As Integer
Visual Basic (Usage)Copy Code
Dim instance As DicomDataSet
Dim value As Integer
 
value = instance.WaveformGroupCount
C# 
public int WaveformGroupCount {get;}
C++/CLI 
public:
property int WaveformGroupCount {
   int get();
}

Property Value

Value that represents the number of waveform multiplex groups that are included in the dataset.

Example

This is a comprehensive sample, which shows how to insert a waveform group with one ECG channel into a dataset. The main method is InsertECGCWaveform ,and the rest of the methods are helping methods.All these methods are assumed to be part of the same class. The main method that creates the waveform group and adds it to the dataset

Visual BasicCopy Code
Public Sub InsertECGCWaveform(ByVal inputDS As DicomDataSet, ByVal samples As Short(), ByVal count As Integer)

   ' Our new waveform group
   Dim ECGWaveformGroup As DicomWaveformGroup = New DicomWaveformGroup()

   ' Reset the waveform group, we don't 
   ' really need to call this! 
   ECGWaveformGroup.Reset()

   ' Set the number of samples per channel. You can call 
   ' DicomWaveformGroup.GetNumberOfSamplesPerChannel to get 
   ' the number of samples
   ECGWaveformGroup.SetNumberOfSamplesPerChannel(count)

   ' Set the sampling frequency. You can call 
   ' DicomWaveformGroup.GetSamplingFrequency to 
   ' get the sampling frequency
   ECGWaveformGroup.SetSamplingFrequency(240.0)

   ' Set sample interpretation
   ECGWaveformGroup.SetSampleInterpretation(DicomWaveformSampleInterpretationType.Signed16BitLinear)

   ' Just in case! 
   Debug.Assert(ECGWaveformGroup.GetSampleInterpretation() = DicomWaveformSampleInterpretationType.Signed16BitLinear)

   ' No Multiplex Group Time Offset
   ' You can call DicomWaveformGroup.GetMultiplexGroupTimeOffset 
   ' to get the time offset
   ECGWaveformGroup.SetMultiplexGroupTimeOffset(False, 0.0)

   ' No Trigger Time Offset
   ' You can call DicomWaveformGroup.GetTriggerTimeOffset 
   ' to get the trigger time offset
   ECGWaveformGroup.SetTriggerTimeOffset(False, 0.0)

   ' No Trigger Sample Position
   ' You can call DicomWaveformGroup.GetTriggerSamplePosition
   ' to get the trigger sample position
   ECGWaveformGroup.SetTriggerSamplePosition(False, 0)

   ' Waveform originality is original   
   ECGWaveformGroup.SetWaveformOriginality(DicomWaveformOriginalityType.Original)

   ' Just in case! 
   Debug.Assert(ECGWaveformGroup.GetWaveformOriginality() = DicomWaveformOriginalityType.Original)

   ' Set the multiplex group label
   ECGWaveformGroup.SetMultiplexGroupLabel("SCPECG Waveform")
   Debug.Assert(ECGWaveformGroup.GetMultiplexGroupLabel() = "SCPECG Waveform")

   ' Set the Waveform padding value
   ' You can call DicomWaveformGroup.GetWaveformPaddingValue
   ' to get the waveform padding value
   ECGWaveformGroup.SetWaveformPaddingValue(True, 32768)

   InsertECGCChannel(ECGWaveformGroup, samples, count)

   ' Delete any waveform groups that already exist in the dataset
   Dim index As Integer = 0
   Do While index < inputDS.WaveformGroupCount
      inputDS.DeleteWaveformGroup(index)
      index += 1
   Loop

   ' Insert the new waveform group into the dataset
   inputDS.AddWaveformGroup(ECGWaveformGroup, 0)
End Sub

' Add an ECG channel to the group
Private Sub InsertECGCChannel(ByVal ECGWaveformGroup As DicomWaveformGroup, ByVal samples As Short(), ByVal count As Integer)

   ' Add a channel to the group
   Dim ECGChannel As DicomWaveformChannel = ECGWaveformGroup.AddChannel(1)
   If ECGChannel Is Nothing Then
      Return
   End If

   ' If we want to update a channel, we can first call 
   ' DicomWaveformGroup.DeleteChannel and then call 
   ' DicomWaveformGroup.AddChannel with the same 
   ' index we passed to DeleteChannel


   ' Make sure that the channel really got added 
   ' This is for the purposes of this sample only, because 
   ' the check we did in the previous statement is enough
   If (ECGWaveformGroup.ChannelCount <> 1) OrElse (ECGWaveformGroup.GetChannel(0) Is Nothing) Then
      Return
   End If

   ' We can call DicomWaveformChannel.GetWaveformGroup to access the waveform
   ' group that the channel belongs to

   If ECGChannel.Index <> 0 Then
      Return
   End If

   '
   'Set the channel samples
   'The data we are setting in here is 16 bit data.
   'We would call SetChannelSamples8 if the data were 8 bit.
   'We can also call SetChannelSamples32, which will set the data
   'as either 8-bit or 16-bit depending on the sample interpretation
   '

   If ECGChannel.SetChannelSamples16(samples, count) <> count Then
      Return
   End If

   If (ECGChannel.GetChannelSamples() Is Nothing) OrElse (ECGChannel.GetChannelSamplesCount() = 0) Then
      Return
   End If

   ' Set the channel source and sensitivity
   SetChannelSourceAndSensitivity(ECGChannel)

   ' Set the channel status
   ECGChannel.SetChannelStatus(DicomChannelStatusType.OK)
   If ECGChannel.GetChannelStatus() <> DicomChannelStatusType.OK Then
      Return
   End If

   ' Set the channel time skew. You can call 
   ' LDicomWaveformChannel::GetChannelTimeSkew to get the time skew
   ' You can also call DicomWaveformChannel.SetChannelSampleSkew
   ' And DicomWaveformChannel.GetChannelSampleSkew to set and get 
   ' the channel samples skew
   ECGChannel.SetChannelTimeSkew(0.0)

   ' Set the waveform channel number
   ECGChannel.SetWaveformChannelNumber(True, 0)
   If ECGChannel.GetWaveformChannelNumber() <> 0 Then
      Return
   End If

   ' Set the channel label
   ECGChannel.SetChannelLabel("First Channel")

   If ECGChannel.GetChannelLabel() <> "First Channel" Then
      Return
   End If

   ' No channel offset
   ' You can also call DicomWaveformChannel.GetChannelOffset
   ' To get the channel offset

   ECGChannel.SetChannelOffset(False, 0.0)

   ' Set filter low frequency
   ' You can also call DicomWaveformChannel.GetFilterLowFrequency
   ' To get the filter low frequency

   ECGChannel.SetFilterLowFrequency(True, 0.05)

   ' Set filter high frequency
   ' You can also call DicomWaveformChannel.GetFilterHighFrequency
   ' To get the filter high frequency

   ECGChannel.SetFilterHighFrequency(True, 100.0)

   ' Set the channel minimum value
   ECGChannel.SetChannelMinimumValue(True, -386)

   ' Set the channel maximum value
   ECGChannel.SetChannelMaximumValue(True, 1264)


   If ECGChannel.GetChannelMinimumValue() <> -386 Then
      Return
   End If

   If ECGChannel.GetChannelMaximumValue() <> 1264 Then
      Return
   End If

   '
   'You can also call the following functions to set and get the notch
   'filter frequency and bandwidth
   'DicomWaveformChannel.SetNotchFilterFrequency
   'DicomWaveformChannel.GetNotchFilterFrequency
   'DicomWaveformChannel.SetNotchFilterBandwidth
   'DicomWaveformChannel.GetNotchFilterBandwidth
   '

   ' Last, but not least, set the channel annotations! 
   SetChannelAnnotations(ECGChannel)
End Sub

'Set the channel source and sensitivity
Private Sub SetChannelSourceAndSensitivity(ByVal channel As DicomWaveformChannel)

   ' Let's use the DICOM context group tables! 

   '/////////////////////Channel Source/////////////////////////
   ' Load the ECG Leads table 

   DicomContextGroupTable.Instance.Load(DicomContextIdentifierType.CID3001)
   Dim ECGLeadsGroup As DicomContextGroup = DicomContextGroupTable.Instance.Find(DicomContextIdentifierType.CID3001)
   If Not ECGLeadsGroup Is Nothing Then
      Dim type As DicomContextIdentifierType = ECGLeadsGroup.ContextIdentifier

      Dim codedConcept As DicomCodedConcept = DicomContextGroupTable.Instance.FindCodedConcept(ECGLeadsGroup, "SCPECG", "5.6.3-9-1")
      If Not codedConcept Is Nothing Then
         Dim codeSequenceItem As DicomCodeSequenceItem = New DicomCodeSequenceItem()
         codeSequenceItem.CodeValue = codedConcept.CodeValue
         codeSequenceItem.CodingSchemeDesignator = codedConcept.CodingSchemeDesignator
         codeSequenceItem.CodingSchemeVersion = codedConcept.CodingSchemeVersion
         codeSequenceItem.CodeMeaning = codedConcept.CodeMeaning

         ' Set the channel source
         channel.SetChannelSource(codeSequenceItem)
         ' You can call DicomWaveformChannel.GetChannelSource to get the channel source
      End If
   End If

   '/////////////////////Channel Sensitivity////////////////////////
   DicomContextGroupTable.Instance.Load(DicomContextIdentifierType.CID3082)
   ' Cardiology Units of Measurement
   Dim cardiologyUnitsGroup As DicomContextGroup = DicomContextGroupTable.Instance.Find(DicomContextIdentifierType.CID3082)
   If Not cardiologyUnitsGroup Is Nothing Then
      Dim codedConcept As DicomCodedConcept = DicomContextGroupTable.Instance.FindCodedConcept(cardiologyUnitsGroup, "UCUM", "mV")
      If Not codedConcept Is Nothing Then
         Dim sourceSequenceItem As DicomCodeSequenceItem = New DicomCodeSequenceItem()

         sourceSequenceItem.CodeValue = codedConcept.CodeValue
         sourceSequenceItem.CodingSchemeDesignator = codedConcept.CodingSchemeDesignator
         sourceSequenceItem.CodingSchemeVersion = codedConcept.CodingSchemeVersion
         sourceSequenceItem.CodeMeaning = codedConcept.CodeMeaning

         ' Set the channel sensitivity
         channel.SetChannelSensitivity(True, 0.00122, sourceSequenceItem, 1.0, 0.0)

         ' You can call DicomWaveformChannel.GetChannelSensitivity to get the channel sensitivity
      End If
   End If
End Sub

' Add annotations for the channel
Private Sub SetChannelAnnotations(ByVal channel As DicomWaveformChannel)

   ' Delete any existing channel annotations
   Dim index As Integer = 0
   Do While index < channel.GetAnnotationCount()
      If Not channel.GetAnnotation(index) Is Nothing Then
         channel.DeleteAnnotation(index)
      End If
      index += 1
   Loop

   Dim waveformAnnotation As DicomWaveformAnnotation = New DicomWaveformAnnotation()
   Dim codedName As DicomCodeSequenceItem = New DicomCodeSequenceItem()

   codedName.CodeValue = "8867-4"
   codedName.CodeMeaning = "Heart rate"
   codedName.CodingSchemeDesignator = "LN"
   codedName.CodingSchemeVersion = "19971101"

   waveformAnnotation.CodedName = codedName

   'Over here we can access other properties such as 
   'waveformAnnotation.CodedValue and waveformAnnotation.UnformattedTextValue 

   Dim measurementUnits As DicomCodeSequenceItem = New DicomCodeSequenceItem()

   measurementUnits.CodeValue = "{H.B.}/min"
   measurementUnits.CodeMeaning = "Heart beat per minute"
   measurementUnits.CodingSchemeDesignator = "UCUM"
   measurementUnits.CodingSchemeVersion = "1.4"

   waveformAnnotation.MeasurementUnits = measurementUnits

   waveformAnnotation.SetNumericValue(New Double() {69.0}, 1)

   channel.AddAnnotation(waveformAnnotation)
End Sub
C#Copy Code
public void InsertECGCWaveform(DicomDataSet inputDS, short[] samples, int count)
{

   // Our new waveform group
   DicomWaveformGroup ECGWaveformGroup = new DicomWaveformGroup();

   // Reset the waveform group, we don't 
   // really need to call this! 
   ECGWaveformGroup.Reset();

   // Set the number of samples per channel. You can call 
   // DicomWaveformGroup.GetNumberOfSamplesPerChannel to get 
   // the number of samples
   ECGWaveformGroup.SetNumberOfSamplesPerChannel(count);

   // Set the sampling frequency. You can call 
   // DicomWaveformGroup.GetSamplingFrequency to 
   // get the sampling frequency
   ECGWaveformGroup.SetSamplingFrequency(240.00);

   // Set sample interpretation
   ECGWaveformGroup.SetSampleInterpretation(DicomWaveformSampleInterpretationType.Signed16BitLinear);

   // Just in case! 
   Debug.Assert(ECGWaveformGroup.GetSampleInterpretation() == DicomWaveformSampleInterpretationType.Signed16BitLinear);

   // No Multiplex Group Time Offset
   // You can call DicomWaveformGroup.GetMultiplexGroupTimeOffset 
   // to get the time offset
   ECGWaveformGroup.SetMultiplexGroupTimeOffset(false, 0.0);

   // No Trigger Time Offset
   // You can call DicomWaveformGroup.GetTriggerTimeOffset 
   // to get the trigger time offset
   ECGWaveformGroup.SetTriggerTimeOffset(false, 0.0);

   // No Trigger Sample Position
   // You can call DicomWaveformGroup.GetTriggerSamplePosition
   // to get the trigger sample position
   ECGWaveformGroup.SetTriggerSamplePosition(false, 0);

   // Waveform originality is original   
   ECGWaveformGroup.SetWaveformOriginality(DicomWaveformOriginalityType.Original);

   // Just in case! 
   Debug.Assert(ECGWaveformGroup.GetWaveformOriginality() == DicomWaveformOriginalityType.Original);

   // Set the multiplex group label
   ECGWaveformGroup.SetMultiplexGroupLabel("SCPECG Waveform");
   Debug.Assert(ECGWaveformGroup.GetMultiplexGroupLabel() == "SCPECG Waveform");

   // Set the Waveform padding value
   // You can call DicomWaveformGroup.GetWaveformPaddingValue
   // to get the waveform padding value
   ECGWaveformGroup.SetWaveformPaddingValue(true, 32768);

   InsertECGCChannel(ECGWaveformGroup, samples, count);

   // Delete any waveform groups that already exist in the dataset
   for (int index = 0; index < inputDS.WaveformGroupCount; index++)
   {
      inputDS.DeleteWaveformGroup(index);
   }

   // Insert the new waveform group into the dataset
   inputDS.AddWaveformGroup(ECGWaveformGroup, 0);
}

// Add an ECG channel to the group
void InsertECGCChannel(DicomWaveformGroup ECGWaveformGroup, short[] samples, int count)
{

   // Add a channel to the group
   DicomWaveformChannel ECGChannel = ECGWaveformGroup.AddChannel(1);
   if (ECGChannel == null)
   {
      return;
   }

   // If we want to update a channel, we can first call 
   // DicomWaveformGroup.DeleteChannel and then call 
   // DicomWaveformGroup.AddChannel with the same 
   // index we passed to DeleteChannel


   // Make sure that the channel really got added 
   // This is for the purposes of this sample only, because 
   // the check we did in the previous statement is enough
   if ((ECGWaveformGroup.ChannelCount != 1) ||
         (ECGWaveformGroup.GetChannel(0) == null))
   {
      return;
   }

   // We can call DicomWaveformChannel.GetWaveformGroup to access the waveform
   // group that the channel belongs to

   if (ECGChannel.Index != 0)
   {
      return;
   }

   /*
      Set the channel samples
      The data we are setting in here is 16 bit data.
      We would call SetChannelSamples8 if the data were 8 bit.
      We can also call SetChannelSamples32, which will set the data
      as either 8-bit or 16-bit depending on the sample interpretation
   */

   if (ECGChannel.SetChannelSamples16(samples, count) != count)
   {
      return;
   }

   if ((ECGChannel.GetChannelSamples() == null) || (ECGChannel.GetChannelSamplesCount() == 0))
      return;

   // Set the channel source and sensitivity
   SetChannelSourceAndSensitivity(ECGChannel);

   // Set the channel status
   ECGChannel.SetChannelStatus(DicomChannelStatusType.OK);
   if (ECGChannel.GetChannelStatus() != DicomChannelStatusType.OK)
   {
      return;
   }

   // Set the channel time skew. You can call 
   // LDicomWaveformChannel::GetChannelTimeSkew to get the time skew
   // You can also call DicomWaveformChannel.SetChannelSampleSkew
   // And DicomWaveformChannel.GetChannelSampleSkew to set and get 
   // the channel samples skew
   ECGChannel.SetChannelTimeSkew(0.000000);

   // Set the waveform channel number
   ECGChannel.SetWaveformChannelNumber(true, 0);
   if (ECGChannel.GetWaveformChannelNumber() != 0)
      return;

   // Set the channel label
   ECGChannel.SetChannelLabel("First Channel");

   if (ECGChannel.GetChannelLabel() != "First Channel")
      return;

   // No channel offset
   // You can also call DicomWaveformChannel.GetChannelOffset
   // To get the channel offset

   ECGChannel.SetChannelOffset(false, 0.0);

   // Set filter low frequency
   // You can also call DicomWaveformChannel.GetFilterLowFrequency
   // To get the filter low frequency

   ECGChannel.SetFilterLowFrequency(true, 0.050);

   // Set filter high frequency
   // You can also call DicomWaveformChannel.GetFilterHighFrequency
   // To get the filter high frequency

   ECGChannel.SetFilterHighFrequency(true, 100.00);

   // Set the channel minimum value
   ECGChannel.SetChannelMinimumValue(true, -386);

   // Set the channel maximum value
   ECGChannel.SetChannelMaximumValue(true, 1264);


   if (ECGChannel.GetChannelMinimumValue() != -386)
      return;

   if (ECGChannel.GetChannelMaximumValue() != 1264)
      return;

   /*
      You can also call the following functions to set and get the notch 
      filter frequency and bandwidth
      DicomWaveformChannel.SetNotchFilterFrequency
      DicomWaveformChannel.GetNotchFilterFrequency
      DicomWaveformChannel.SetNotchFilterBandwidth
      DicomWaveformChannel.GetNotchFilterBandwidth
   */

   // Last, but not least, set the channel annotations! 
   SetChannelAnnotations(ECGChannel);
}

//Set the channel source and sensitivity
void SetChannelSourceAndSensitivity(DicomWaveformChannel channel)
{

   // Let's use the DICOM context group tables! 

   ///////////////////////Channel Source/////////////////////////
   // Load the ECG Leads table 

   DicomContextGroupTable.Instance.Load(DicomContextIdentifierType.CID3001);
   DicomContextGroup ECGLeadsGroup = DicomContextGroupTable.Instance.Find(DicomContextIdentifierType.CID3001);
   if (ECGLeadsGroup != null)
   {
      DicomContextIdentifierType type = ECGLeadsGroup.ContextIdentifier;

      DicomCodedConcept codedConcept = DicomContextGroupTable.Instance.FindCodedConcept(ECGLeadsGroup, "SCPECG", "5.6.3-9-1");
      if (codedConcept != null)
      {
         DicomCodeSequenceItem codeSequenceItem = new DicomCodeSequenceItem();
         codeSequenceItem.CodeValue = codedConcept.CodeValue;
         codeSequenceItem.CodingSchemeDesignator = codedConcept.CodingSchemeDesignator;
         codeSequenceItem.CodingSchemeVersion = codedConcept.CodingSchemeVersion;
         codeSequenceItem.CodeMeaning = codedConcept.CodeMeaning;

         // Set the channel source
         channel.SetChannelSource(codeSequenceItem);
         // You can call DicomWaveformChannel.GetChannelSource to get the channel source
      }
   }

   ///////////////////////Channel Sensitivity////////////////////////
   DicomContextGroupTable.Instance.Load(DicomContextIdentifierType.CID3082);
   // Cardiology Units of Measurement
   DicomContextGroup cardiologyUnitsGroup = DicomContextGroupTable.Instance.Find(DicomContextIdentifierType.CID3082);
   if (cardiologyUnitsGroup != null)
   {
      DicomCodedConcept codedConcept = DicomContextGroupTable.Instance.FindCodedConcept(cardiologyUnitsGroup, "UCUM", "mV");
      if (codedConcept != null)
      {
         DicomCodeSequenceItem sourceSequenceItem = new DicomCodeSequenceItem();

         sourceSequenceItem.CodeValue = codedConcept.CodeValue;
         sourceSequenceItem.CodingSchemeDesignator = codedConcept.CodingSchemeDesignator;
         sourceSequenceItem.CodingSchemeVersion = codedConcept.CodingSchemeVersion;
         sourceSequenceItem.CodeMeaning = codedConcept.CodeMeaning;

         // Set the channel sensitivity
         channel.SetChannelSensitivity(true, 0.001220, sourceSequenceItem, 1.0, 0.0);

         // You can call DicomWaveformChannel.GetChannelSensitivity to get the channel sensitivity
      }
   }
}

// Add annotations for the channel
void SetChannelAnnotations(DicomWaveformChannel channel)
{

   // Delete any existing channel annotations
   for (int index = 0; index < channel.GetAnnotationCount(); index++)
   {
      if (channel.GetAnnotation(index) != null)
      {
         channel.DeleteAnnotation(index);
      }
   }

   DicomWaveformAnnotation waveformAnnotation = new DicomWaveformAnnotation();
   DicomCodeSequenceItem codedName = new DicomCodeSequenceItem();

   codedName.CodeValue = "8867-4";
   codedName.CodeMeaning = "Heart rate";
   codedName.CodingSchemeDesignator = "LN";
   codedName.CodingSchemeVersion = "19971101";

   waveformAnnotation.CodedName = codedName;
   //Over here we can access other properties such as 
   // waveformAnnotation.CodedValue and waveformAnnotation.UnformattedTextValue 

   DicomCodeSequenceItem measurementUnits = new DicomCodeSequenceItem();

   measurementUnits.CodeValue = "{H.B.}/min";
   measurementUnits.CodeMeaning = "Heart beat per minute";
   measurementUnits.CodingSchemeDesignator = "UCUM";
   measurementUnits.CodingSchemeVersion = "1.4";

   waveformAnnotation.MeasurementUnits = measurementUnits;

   waveformAnnotation.SetNumericValue(new double[] { 69.00 }, 1);

   channel.AddAnnotation(waveformAnnotation);
}
SilverlightCSharpCopy Code
SilverlightVBCopy Code

Remarks

In DICOM, a waveform consists of one or more multiplex groups, each encoded into an item in the "Waveform Sequence" (5400,0100). Use this method to retrieve the number of items under that sequence.

Once you know the total number of multiplex groups, call the Leadtools.Dicom.DicomDataSet.GetWaveformGroup method to extract a specific group or the Leadtools.Dicom.DicomDataSet.DeleteWaveformGroup method to delete it.

Requirements

Target Platforms: Silverlight 3.0, Windows XP, Windows Server 2003 family, Windows Server 2008 family, Windows Vista, Windows 7, MAC OS/X (Intel Only)

See Also

Leadtools.Dicom requires a Medical toolkit server license and unlock key. For more information, refer to: Imaging Pro/Document/Medical Features