StentEnhancementCommand Class

Summary

Performs a digital stent enhancement on a sequence of X-Ray frames. This enhances the visibility of the stent device in X-Ray angiography sequence of the coronary arteries.

Remarks
  • This command enhances the visibility of the stent device by compensating for motion and integrating the frames of the X-Ray sequence.
  • The output of the command is an averaged image for which motion has been compensated.
  • To use the command:

  • Load the sequence of frames to be enhanced.
  • Choose a reference frame and draw a rectangle which covers the stent device throughout all of the frames.
  • To detect the markers for the stent device, either use the DetectMarkers method, or input them using the FirstMarker and SecondMarker properties.
  • Use the RunCommand to perform the enhancement process.
  • Obtain the averaged image from the EnhancedImage property, then apply different contrast enhancements, such as the MultiscaleEnhancementCommand.
  • This command does not support camera motion (C-Arm motion).

  • To get the best result, the sequence should contain more than 40 frames.
  • This command supports 8, 12, 16-bit grayscale images.